• DrugWipe® F
Instructions for UseTest for surfaces to detect: Cannabis, Opiates, Cocaine and Amphetamines/Methamphetamines
  • Discard the test if the expiration date has been exceeded.
  • Remove the red sample collector from the white test cassette. Do not remove the protective cap. Do not touch the sampling pads.
  • Moisten the sampling pads on the sample collector. Shake off any excess water.
  • Using slight pressure, wipe surface with the sample collector approx. 5 to 6 times. Only dab the surface if it is very rough.
  • Place the sample collector back on the test cassette. The sample collector must lock in place with an audible double click.
  • Remove the protective cap and fil with water up to the marked level. Immerse the test strip in water for 15 seconds and then remove. Do not immerse the white plastic casing when doing so.
  • Leave the test undisturbed on a horizontal surface. After 8 minutes, read the test result.
  • Results interpretation